Jun 08, 2012 Download the Agilent G1888 Network Headspace Sampler user manual,or search and download other millions of manuals/guides/handbooks from ManualLib. Buy and sell Gas Chromatography Samplers and Autosamplers using auctions and classifieds. New, Surplus, and Used GC Autosamplers for sale. Headspace sampling allows you to introduce volatile compounds automatically, from virtually any sample matrix, directly into a GC or GC/MS. The Agilent G1888 network headspace sampler features an inert sample pathway for superior chemical performance without analyte degradation or loss. Here we transferred a blood alcohol analysis method from Agilent’s G1888A Headspace Sampler to the 7697A Headspace Sampler. Agilent’s 7697A Headspace Sampler also gives you an added benefit: shorter cycle times for blood alcohols without sacrificing resolution. The sampler’s proven architecture, together with Agilent’s exclusive sampling.

Agilent G1888 Headspace Autosampler Manual
I'm thinking the only thing it could be is the 6-port valve is contaminated because that's the only thing that's a variable when opening the sample valve besides the sample loop. When I did call Agilent last time I asked if steam cleaning could fix the contamination and they said that rarely works. I tried calling them today but they were closed. I did look the 6-port valve up and it's over $1000 so I don't want to waste the money if there's something else I could try first.
Oh, one more thing. We typically run around 90 oven, 110 loop and 115 transfer line. I cranked all three to 250 and I was getting a lot more response then I was previously. Another test I did was watched the output for a while, which was around 15, and then opened the sample valve. It would jump up to around 250-300 and then settle back down to around 20, constantly. It would sit at 20 forever until I closed the sample valve and then it would drop back down to 15, showing that flushing through the 6-port valve and sample loop was causing static contamination to be sent through the system.
Thanks in advance for any help!